2.4 Mounting
The Model 231RS can be easily mounted using the two mounting screws located on the side of the unit.
2.5 Installation Procedures
For dierential pressure measurements at high line pressure, it is recommended that the pressure sensor be installed
with a valve in each line, plus a shunt valve across the high and low (reference) pressure ports as shown.
Important: Do not exceed maximum range pressure with the total of dierential pressure and line pressure.
Valve A
1
High
Low Pressure
Remote Sensor
High Pressure
Remote Sensor
Valve B
1
Low
Model 231RS
Valve A = High Side Valve
Valve B = Low Side Valve
1
Valves not included.
PipePipe
Figure 1